Miwa H, Matsuda T, Tani N, Kondo K, Iida F
Department of Biomedical Engineering, Research Institute, Osaka, Japan.
ASAIO J. 1993 Jul-Sep;39(3):M501-5.
The authors developed a hybrid vascular graft consisting of three components: a polyurethane (PU) graft (internal diameter = 3 mm; length: 5 cm) with a compliance similar to that of natural arteries; an artificial basement membrane composed of a complex gel of Type I collagen and dermatan sulfate; and an autogenous endothelial cell (EC) monolayer constructed before implantation. As a continuation of their previous study that focused on antithrombogenicity for as long as 1 month after implantation, the authors evaluated the antihyperplastic characteristics of their hybrid vascular graft for longer implantation periods. The hybrid grafts were implanted bilaterally into carotid arteries of dogs without anticoagulant or anti-platelet agents. The implantation period ranged from 6 to 26 weeks. The overall patency rate observed in this study was 70%. Histologic evaluation of grafts revealed that luminal surfaces of the arteries and grafts joined uniformly and smoothly. The mean initimal thickness of grafts was approximately 80 microns at 12 weeks; there was no statistically significant difference in the thickness between portions of the graft. Little additional increase was observed at 26 weeks. The intimal thickness of adjacent arteries was small. Thus, anastomotic hyperplasia was considered minimized. These encouraging results were explained in terms of synergistic activities of the compliant PU graft and durable EC monolayer as follows. The PU graft, which was fabricated to have the same compliance as that of natural arteries, minimized the compliance mismatch at the anastomoses.(ABSTRACT TRUNCATED AT 250 WORDS)
一种内径为3毫米、长度为5厘米且顺应性与天然动脉相似的聚氨酯(PU)移植物;一种由I型胶原蛋白和硫酸皮肤素的复合凝胶组成的人工基底膜;以及在植入前构建的自体内皮细胞(EC)单层。作为其先前专注于植入后长达1个月抗血栓形成研究的延续,作者评估了他们的混合血管移植物在更长植入期的抗增生特性。将混合移植物双侧植入未使用抗凝剂或抗血小板药物的犬颈动脉。植入期为6至26周。本研究中观察到的总体通畅率为70%。对移植物的组织学评估显示,动脉和移植物的管腔表面均匀且平滑地连接。在12周时移植物的平均内膜厚度约为80微米;移植物各部分之间的厚度无统计学显著差异。在26周时观察到内膜厚度几乎没有进一步增加。相邻动脉的内膜厚度较小。因此,吻合口增生被认为降至最低。这些令人鼓舞的结果可以通过顺应性PU移植物和持久的EC单层的协同作用来解释如下。制造的PU移植物具有与天然动脉相同的顺应性,使吻合处的顺应性不匹配最小化。