Variations in acoustical beam properties of intracoronary Doppler catheters.

作者信息

Roth T, Erbel R, Brennecke R, Rupprecht H J, Meyer J, von Seelen W

机构信息

2nd Medical Clinic, Johannes Gutenberg University, Mainz, Germany.

出版信息

Cathet Cardiovasc Diagn. 1993 Nov;30(3):257-63. doi: 10.1002/ccd.1810300318.

Abstract

The limitations of coronary angiography in assessing the functional significance of coronary obstructions is well known. While the critical variable of coronary blood flow cannot be readily measured, intraluminal Doppler sonography offers useful related functional information on blood flow velocity. In order to fully evaluate Doppler signals it is essential to have exact knowledge of the transducer transmission characteristics and of the ultrasound beam topology. In an experimental set-up, the transmitter-receiver characteristics of five commonly used Doppler catheters were investigated. In comparing the beam characteristics we found inhomogeneities in the lateral beam spread. At a penetration depth of 3.0 mm the beam shape varied from a minimum of 1.25 mm up to a maximum of 3.5 mm. The mean was 2.25 mm. The different beam profiles of the investigated Doppler transducers cause an error in measuring the blood flow velocity. The blood flow velocity tends to be underestimated the more the vessel diameter and the blood flow velocity increase. Contrary to transducer design optimized for imaging, for spectral analysis of the Doppler signal it would be advantageous to have as broad a beam as possible in order to illuminate the entire vessel lumen.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验