Separation of DNA fragments by capillary electrophoresis using replaceable linear polyacrylamide matrices.

作者信息

Pariat Y F, Berka J, Heiger D N, Schmitt T, Vilenchik M, Cohen A S, Foret F, Karger B L

机构信息

Barnett Institute, Northeastern University, Boston, MA 02115.

出版信息

J Chromatogr A. 1993 Oct 15;652(1):57-66. doi: 10.1016/0021-9673(93)80645-O.

Abstract

The use of low percent (1.5-6% T) replaceable linear polyacrylamide (LPA) network matrices for rapid separation of double-stranded DNA fragments was explored. Separations of fragments ranging from 20 to 23,000 base pairs were readily achieved. Typically, 4 x 10(6) theoretical plates/m were obtained in less than 30 min. Short separation times under 2 min were also possible, using the DNA intercalating dye, ethidium bromide, along with high electric fields. The high resolving power of linear polyacrylamide was demonstrated in the separation of two fragments which differ by a single base pair (123/124 base pairs) using 6% T LPA and ethidium bromide intercalation. This LPA composition allowed for the possible single base-pair resolution of dsDNA fragments up to 300 base pairs in length. Several concentrations of the linear polyacrylamide for different ranges of fragment lengths have been employed. In addition, replaceable LPA offers the advantage of a fresh separation matrix for each run, thus overcoming column stability problems and minimizing needs for sample cleanup. Electro-osmotic flow was substantially reduced using stable capillary coatings, which were required for obtaining high efficiencies and good reproducibility.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验