局部类固醇洗脱对心内膜起搏器导线慢性起搏阈值和阻抗的影响。

Effects of local steroid elution on chronic pacing threshold and impedance of endocardial pacemaker leads.

作者信息

Schuchert A, Kuck K H

机构信息

Department of Cardiology, University Hospital Eppendorf, Hamburg, FRG.

出版信息

Cardiology. 1993;83(4):240-3. doi: 10.1159/000175976.

Abstract

To study the effects of steroid elution on chronic pacing threshold and impedance, a 4-mm2 surface area lead with steroid (030-368, Telectronics; n = 9) was compared to the same lead without steroid (030-359, Telectronics; n = 9) 3 months after implantation. The pulse duration thresholds of the steroid lead were significantly lower at 0.8, 1.6, 2.5, and 5.0 V. Impedance measured by telemetry at 0.8, 1.6, 2.5, 5.0 and 8.0 V pulse amplitude, and 0.05, 0.25, 0.5 and 1.0 ms pulse duration was similar for the two leads. Mean impedance was about 650 omega and did not change from 0.8 to 8.0 V. The local steroid elution reduced the pacing threshold, but had no influence on the impedance. The stable impedance indicates that the impedance of this lead was not modified by pulse-dependent variables and, therefore, can also be correctly determined at pulse amplitudes below 5.0 V.

摘要

为研究类固醇洗脱对慢性起搏阈值和阻抗的影响,将植入3个月后带有类固醇的4平方毫米表面积导线(030 - 368,泰利特隆尼克斯公司;n = 9)与相同但无类固醇的导线(030 - 359,泰利特隆尼克斯公司;n = 9)进行比较。在0.8、1.6、2.5和5.0伏时,含类固醇导线的脉冲持续时间阈值显著更低。通过遥测在0.8、1.6、2.5、5.0和8.0伏脉冲幅度以及0.05、0.25、0.5和1.0毫秒脉冲持续时间下测量的阻抗,两根导线相似。平均阻抗约为650欧姆,从0.8伏到8.0伏未发生变化。局部类固醇洗脱降低了起搏阈值,但对阻抗无影响。稳定的阻抗表明该导线的阻抗不受脉冲相关变量的影响,因此,在低于5.0伏的脉冲幅度下也能正确测定。