Tartaglino L M, Flanders A E, Vinitski S, Friedman D P
Department of Radiology, Jefferson Medical College, Philadelphia, PA.
Radiology. 1994 Feb;190(2):565-9. doi: 10.1148/radiology.190.2.8284417.
To determine whether the relative insensitivity of T2-weighted fast spin-echo (FSE) techniques to magnetic susceptibility can be exploited to reduce metallic artifacts on images of the postoperative spine and, thus, improve the interpretation of the postoperative study.
Three neuroradiologists retrospectively evaluated sagittal T2-weighted conventional spin-echo and FSE images obtained in 15 patients with metallic artifacts from various sources including drill particles from anterior cervical diskectomy, posterior fixation wires, fixation rods or plates, and an inferior vena cava filter. The amount of artifact present and whether these artifacts affected image interpretation were evaluated.
Among the 45 paired evaluations, the artifact was judged to be less apparent with FSE sequences in 39. In eight of 45 evaluations (18%), the interpretation of the area of interest was possible only on the FSE images.
FSE imaging, especially when performed with shorter echo spacing, increases the amount of T2-weighted information in the presence of metallic artifact because it decreases magnetic susceptibility effects.
确定是否可以利用T2加权快速自旋回波(FSE)技术对磁化率相对不敏感的特性来减少术后脊柱图像上的金属伪影,从而改善术后检查的解读。
三位神经放射科医生回顾性评估了15例存在金属伪影患者的矢状面T2加权常规自旋回波和FSE图像,这些金属伪影来源各异,包括颈椎前路椎间盘切除术后的钻屑、后路固定钢丝、固定棒或钢板以及下腔静脉滤器。评估了存在的伪影量以及这些伪影是否影响图像解读。
在45次配对评估中,39次判定FSE序列的伪影不太明显。在45次评估中的8次(18%),仅在FSE图像上才能对感兴趣区域进行解读。
FSE成像,尤其是采用较短回波间隔进行成像时,由于减少了磁化率效应,在存在金属伪影的情况下会增加T2加权信息的量。