Park H J, Lee S Y, Choi S H, Hong H P, Choi Y J, Kim M S
Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Department of Radiology, Kangbuk Samsung Hospital, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
Clin Radiol. 2017 Nov;72(11):996.e1-996.e6. doi: 10.1016/j.crad.2017.07.012. Epub 2017 Aug 12.
To compare multi-point Dixon magnetic resonance imaging (MRI) and fast spin-echo (FSE) T2-weighted imaging (WI) with regard to the size of metallic artefacts when imaging the knee joint.
A total of 42 patients who underwent anterior cruciate ligament (ACL) reconstruction and follow-up imaging with 3 T MRI using the multi-point Dixon technique was included in this retrospective study. The maximal distance of the image distortion area around the metallic artefact was measured (interference screw of femoral tunnel area) on sagittal images of both FSE T2WI (T2WI and fat-suppressed [FS] T2WI) and multi-point Dixon (water only image and in-phase image) sequences. The maximal distance of the image distortion were compared using paired t-tests across the image sequences (multi-point Dixon water only image versus FS T2WI and multi-point Dixon in-phase image versus T2WI).
The mean distance of the image distortion from metallic artefacts regardless of the image sequence ranged from 16.6 mm to 24.5 mm (Table 2). The mean distances measured by two readers on multi-point Dixon (water only image) sequences were significantly shorter than those in FS T2WI sequences (p<0.001). In contrast, the mean distances measured by two readers on multi-point Dixon (in-phase image) sequences did not differ from those of T2WI (p>0.05) sequences.
The water-only image of multi-point Dixon technique reduces the amount of metallic artefacts compared to that in FS FSE T2WI sequences; however, the metallic artefacts were not significantly different between in-phase images of multi-point Dixon and FSE T2WI.
比较多点狄克逊磁共振成像(MRI)和快速自旋回波(FSE)T2加权成像(WI)在膝关节成像时金属伪影的大小。
本回顾性研究纳入了42例行前交叉韧带(ACL)重建并使用3T MRI的多点狄克逊技术进行随访成像的患者。在FSE T2WI(T2WI和脂肪抑制[FS]T2WI)序列以及多点狄克逊序列(仅水图像和同相位图像)的矢状位图像上,测量金属伪影周围图像失真区域的最大距离(股骨隧道区域的干扰螺钉)。使用配对t检验比较各图像序列间图像失真的最大距离(多点狄克逊仅水图像与FS T2WI,以及多点狄克逊同相位图像与T2WI)。
无论图像序列如何,金属伪影导致的图像失真平均距离在16.6毫米至24.5毫米之间(表2)。两位读者在多点狄克逊(仅水图像)序列上测量的平均距离明显短于FS T2WI序列(p<0.001)。相比之下,两位读者在多点狄克逊(同相位图像)序列上测量的平均距离与T2WI序列无差异(p>0.05)。
与FS FSE T2WI序列相比,多点狄克逊技术的仅水图像减少了金属伪影的量;然而,多点狄克逊同相位图像和FSE T2WI之间的金属伪影没有显著差异。