Huysmans M C, Peters M C, Van der Varst P G, Plasschaert A J
TRIKON, Institute for Dental Clinical Research, Dental School, University of Nijmegen, The Netherlands.
Int Endod J. 1993 Sep;26(5):294-300. doi: 10.1111/j.1365-2591.1993.tb00574.x.
Evaluation of the long-term behaviour of restorations in clinical trials can be time-consuming. A partial alternative to the clinical trial can be found in mechanical fatigue testing. The aim of this study was to evaluate the failure behaviour of post and core restored teeth when subjected to cyclic mechanical loading and to compare it with quasistatic failure. Eighty seven premolar teeth were restored with a titanium alloy post and an amalgam or composite core. Five to 21 days after restoration, the specimens were subjected to cyclic loading (frequency 5 Hz), at an angle of 45 degrees to the long axis of the tooth. The load levels were 50, 60, 65 and 70% of mean quasistatic failure loads. The specimens were divided into three groups according to their survival time: short (S) (< 10(4) cycles), intermediate (I) (10(4) < or = life < 10(5) cycles) and long (L) (> or = 10(5) cycles). For both core materials failure behaviour changed after approximately 10(5) cycles, and the change was most marked for the composite group. Catastrophic fatigue failure consisted of core fracture in the amalgam group (three times) and of post fracture in the composite L group (four times). Three post fractures occurred at a site theoretically predisposed to fatigue failure. It was concluded that fatigue failure characteristics of post and core restorations may be very different from those of quasistatic failure. Therefore, in addition to quasistatic tests, fatigue tests are necessary, covering at least 10(5) load cycles.