Kitamura A, Yamada K, Ishihara Y, Ogawa R
Department of Anesthesiology, Nippon Medical School, Tokyo.
Masui. 1993 Jun;42(6):826-30.
During inhalation anesthesia using N2O, the intracuff pressure has been reported to increase due to diffusion of N2O into the cuff. The excessive intracuff pressure may produce ischemia of the tracheal mucosa. In this study, the changes in the intracuff pressure of endotracheal tube with a cuff made of a material with high N2O gas-barrier properties were compared with those of the normal endotracheal tubes. Forty adult patients (16 males and 24 females) were anesthetized with oxygen 2 l.min-1, N2O 4 l.min-1 and isoflurane under controlled ventilation and they were divided into the following three groups according to the type of the endotracheal tube used: the tube with a standard cuff made by PORTEX Co. Ltd; the tube with a profile cuff (high volume cuff) made by PORTEX Co. Ltd; the tube with gas-barrier cuff made by TERUMO Co. Ltd. The changes of each intracuff pressure were recorded continuously for 180 minutes after cuff inflation. A marked increase of the pressure was observed after inhalation of N2O in standard and profile cuff groups. In gas-barrier cuff group, there was a tendency of increasing cuff pressure without a statistically significant difference. The present study suggests that the use of a cuff with a material of high N2O gas-barrier properties would be effective to prevent increased intracuff pressure by N2O diffusion, and this would be especially useful during long anesthesia and hypotensive anesthesia.
据报道,在使用一氧化二氮进行吸入麻醉期间,由于一氧化二氮扩散至气管导管套囊内,套囊内压力会升高。套囊内压力过高可能导致气管黏膜缺血。在本研究中,将由具有高一氧化二氮气体阻隔性能材料制成套囊的气管导管的套囊内压力变化,与普通气管导管的进行了比较。40例成年患者(16例男性和24例女性)在控制通气下接受2 l·min⁻¹氧气、4 l·min⁻¹一氧化二氮和异氟烷麻醉,并根据所用气管导管类型分为以下三组:由PORTEX有限公司生产的带有标准套囊的导管;由PORTEX有限公司生产的带有异形套囊(大容量套囊)的导管;由TERUMO有限公司生产的带有气体阻隔套囊的导管。套囊充气后连续180分钟记录各套囊内压力的变化。在标准套囊组和异形套囊组中,吸入一氧化二氮后观察到压力显著升高。在气体阻隔套囊组中,套囊压力有升高趋势,但无统计学显著差异。本研究表明,使用具有高一氧化二氮气体阻隔性能材料的套囊可有效防止因一氧化二氮扩散导致的套囊内压力升高,这在长时间麻醉和低血压麻醉期间尤其有用。