Between April 1991 and February 1992 we obtained 87 percutaneous core needle biopsy specimens from 60 patients with the automatic biopsy gun. In 78 cases (89.7%) enough renal tissue was obtained for histological analysis. Acute rejection was present on 35 biopsies (40.3%) indicating antirejection therapy. No pathological signs of rejection were detected on 43 biopsies (49.4%). In 38 instances (43.7%) biopsy diagnosis differed from clinical presumptive diagnosis, affecting patient management, while in the remainder it was helpful to confirm clinical impressions. The only complication of the procedure was gross hematuria (requiring vesical irrigation in 1 patient). Due to its simplicity and low morbidity rate, renal biopsy with the automated needle system is the procedure of choice for evaluation and management of renal allograft dysfunction.

摘要

1991年4月至1992年2月期间,我们使用自动活检枪从60例患者身上获取了87份经皮芯针活检标本。78例(89.7%)获取了足够的肾组织用于组织学分析。35份活检标本(40.3%)存在急性排斥反应,提示需进行抗排斥治疗。43份活检标本(49.4%)未检测到排斥反应的病理迹象。38例(43.7%)活检诊断与临床初步诊断不同,影响了患者的治疗管理,而在其余病例中,活检有助于证实临床印象。该操作唯一的并发症是肉眼血尿(1例患者需要膀胱冲洗)。由于其操作简单且发病率低,使用自动针系统进行肾活检是评估和管理肾移植功能障碍的首选方法。