Romaniuk P, Thieme T, Miersch G, Stobbe C, Stösslein F
Universitätsklinikum Charité Berlin, Institut für kardiovaskuläre Diagnostik.
Z Kardiol. 1993;82 Suppl 2:35-40.
The long-term results of 222 patients (pts) with transvenously implanted vena cava filters are critically reviewed. A total of 98 pts were given a Mobin-Uddin filter, 102 pts a Günther filter, and 16 pts a LGM filter. Among the serious complications observed during long-term follow-up were vena cava occlusions and increasing symptoms ranging from congestion of the deep leg veins, perforations into ureters, fractures and embolizations of the filters in up to 40% of the patients, if Günther filters were implanted.
对222例经静脉植入腔静脉滤器的患者的长期结果进行了严格审查。共有98例患者植入了莫宾-乌丁滤器,102例患者植入了君特滤器,16例患者植入了LGM滤器。在长期随访期间观察到的严重并发症包括腔静脉阻塞以及症状加重,症状范围从下肢深静脉充血、输尿管穿孔、滤器骨折和栓塞,若植入君特滤器,高达40%的患者会出现这些情况。