Twardowski Z J, Van Stone J C, Jones M E, Klusmeyer M E, Haynie J D
Department of Medicine, University of Missouri, Columbia 65212.
J Am Soc Nephrol. 1993 Jun;3(12):1978-81. doi: 10.1681/ASN.V3121978.
Long-term i.v. catheters for hemodialysis have the outflow tip extending approximately 2 to 3 cm beyond the inflow tip to prevent blood recirculation during dialysis; however, the lumens are frequently reversed because of inflow failure (i.e., inadequate flow when the inflow lumen is used for blood inflow into the dialyzer). Blood recirculation with reversed lumens (outflow lumen used for blood inflow) in inflow failure catheters and with standard and reversed lumens in well-functioning catheters was measured. Recirculation was measured at a blood flow of 300 mL/min. Systemic blood samples were taken after blood flow was slowed to 50 mL/min. Blood recirculation was calculated as a percentage of the difference between systemic and inflow lumen solute concentrations divided by the difference between systemic and outflow lumen solute concentrations. For each catheter, the recirculation values were calculated separately for urea and creatinine. Average recirculation as measured by both solutes was also calculated. Blood recirculations with standard lumens of well-functioning catheters, reversed lumens of well-functioning catheters, and reversed lumens of inflow failure catheters were 2.09 +/- 1.95, 13.58 +/- 9.87, and 7.10 +/- 5.12 (mean +/- SD), respectively. Whereas recirculation with standard lumens of well-functioning catheters is negligible, reversal of lumens causes considerable recirculation. Recirculation in inflow failure catheters with reversed lumens is significantly less than that with reversed lumens of well-functioning catheters. It was proposed that a blood clot attached at and/or immediately distal to the inflow lumen may disperse outflowing blood and diminish recirculation in inflow failure catheters.(ABSTRACT TRUNCATED AT 250 WORDS)
用于血液透析的长期静脉导管,其流出端比流入端长约2至3厘米,以防止透析期间血液再循环;然而,由于流入故障(即当流入腔用于将血液流入透析器时流量不足),管腔经常会颠倒。测量了流入故障导管中管腔颠倒(流出腔用于血液流入)时以及功能良好的导管中标准管腔和颠倒管腔时的血液再循环情况。在血流速度为300毫升/分钟时测量再循环情况。血流速度减慢至50毫升/分钟后采集全身血样。血液再循环计算为全身与流入腔溶质浓度之差除以全身与流出腔溶质浓度之差的百分比。对于每根导管,分别计算尿素和肌酐的再循环值。还计算了两种溶质测量的平均再循环情况。功能良好的导管标准管腔、功能良好的导管颠倒管腔以及流入故障导管颠倒管腔的血液再循环分别为2.09±1.95、13.58±9.87和7.10±5.12(平均值±标准差)。功能良好的导管标准管腔的再循环可忽略不计,而管腔颠倒会导致相当程度的再循环。流入故障导管中管腔颠倒时的再循环明显低于功能良好的导管中管腔颠倒时的再循环。有人提出,附着在流入腔及其紧邻远端的血凝块可能会分散流出的血液,并减少流入故障导管中的再循环。(摘要截取自250字)