超声仪器性能调查及常规监测的可行性

Performance survey of ultrasound instrumentation and feasibility of routine monitoring.

作者信息

Chirstensen S L, Carson P L

出版信息

Radiology. 1977 Feb;122(2):449-54. doi: 10.1148/122.2.449.

Abstract

Short quality control tests for ultrasound instruments were developed based on the AIUM standard 100mm test object and procedures recommended for its use. While demonstrating the tests of technologists on their instruments, the authors obtained performance data on instruments in 22 hospitals. Subsequent quality control testing on the same instruments demonstrated that manufacturer's servicemen usually had improved performance significantly in the rather large number of cases in which adjustments were indicated by initial tests. Technologists in several hospitals, as well as the authors, performed the tests routinely for extended periods of time, with a moderately high yield for detection or demonstration of malfunctions.

摘要

基于美国医学超声学会(AIUM)标准的100毫米测试物体及其推荐使用程序,开发了超声仪器的简短质量控制测试。在向技术人员演示其仪器的测试时,作者获得了22家医院仪器的性能数据。随后对同一仪器进行的质量控制测试表明,在初始测试表明需要进行调整的相当多案例中,制造商的维修人员通常能显著提高仪器性能。几家医院的技术人员以及作者长期定期进行这些测试,对故障的检测或显示有一定的高成功率。