Bubriak O A, Soldatkin A P, Starodub N F, El'skaia A V, Shul'ga A A, Strikha V I
Ukr Biokhim Zh (1978). 1993 Jan-Feb;65(1):110-3.
The article deals with the optimal conditions of urea determination in solution by the enzyme sensor based on pH-sensitive field effect transistor. The dependence of response level on temperature, pH, ionic strength, buffer capacity were investigated. 10 mM phosphate buffer, pH 7.4 containing 200 mM NaCl is optimal for possible measurements of urea in diluted blood. Temperature change in the measured samples from 15 degrees to 30 degrees C does not affect the sensor signal value. Stability of immobilized ureas was investigated at storage under different conditions. In the presence of 1 mM EDTA activity of immobilized enzyme does not change during one month storage at 4 degrees C.
本文探讨了基于pH敏感场效应晶体管的酶传感器测定溶液中尿素的最佳条件。研究了响应水平与温度、pH值、离子强度、缓冲容量的关系。含200 mM氯化钠的10 mM pH 7.4磷酸盐缓冲液最适合用于稀释血液中尿素的可能测量。被测样品温度从15摄氏度变化到30摄氏度不影响传感器信号值。研究了固定化尿素在不同条件下储存时的稳定性。在1 mM EDTA存在下,固定化酶在4摄氏度储存一个月期间活性不变。