Suppr超能文献

Room-temperature thermodilution cardiac output: proximal injectate lumen vs proximal infusion lumen.

作者信息

Pesola G R, Ayala B, Plante L

机构信息

Department of Anesthesiology and Critical Care Medicine, Albert Einstein University at Montefiore Hospital, Bronx, NY.

出版信息

Am J Crit Care. 1993 Mar;2(2):132-3.

PMID:8358461
Abstract

OBJECTIVE

To assess the accuracy of room-temperature thermodilution cardiac output measurements from the venous infusion port.

DESIGN

Central venous port cardiac output measurements were compared with venous infusion port measurements in 48 right-heart catheters.

INTERVENTION

Three 10-mL injections of 5% dextrose in water were made through each port. The order of port injection was random.

RESULTS

The cardiac outputs were 5.8 +/- 1.8 L/min from both ports, with no difference between ports (paired t test).

CONCLUSION

Room-temperature thermodilution cardiac output determinations from the venous infusion port can be used in place of central venous port cardiac outputs if the central venous port becomes nonfunctional.