Cywinski J K, Hahn A W, Nichols M F, Easley J R
Pacing Clin Electrophysiol. 1978 Jan;1(1):117-25. doi: 10.1111/j.1540-8159.1978.tb03448.x.
Seventeen unipolar cardiac pacemakers powered by hybrid biogalvanic cells were implanted in dogs. Long term clinical effects and electrical performance in vivo of the generators were investigated. The biogalvanic cells were designed to provide 50 years of generator operation. No systemic pathological changes which would be attributed to the generator or the biogalvanic cell were observed. Local reaction to implants was mild and not significantly different from the one observed in conventional pacemaker implantations to date. The electrical performance of the biogalvanic cells was very encouraging. Final steady-state cell voltage levels of .65V to .75V were reached in 100--150 days after implantation and remained constant in 3 units to date which represents over 36 months after implantation for each generator.
将17个由混合生物燃料电池供电的单极心脏起搏器植入犬体内。研究了该起搏器在体内的长期临床效果和电性能。这些生物燃料电池设计可支持起搏器运行50年。未观察到可归因于起搏器或生物燃料电池的全身性病理变化。植入部位的局部反应轻微,与迄今为止传统起搏器植入时观察到的反应无显著差异。生物燃料电池的电性能非常令人鼓舞。植入后100至150天达到最终稳态电池电压水平0.65V至0.75V,迄今为止,有3个装置的该电压水平保持恒定,每个起搏器植入后已超过36个月。