Mugica J, Duconge R, Dubos M, Lazarus B
Pacing Clin Electrophysiol. 1978 Apr;1(2):222-30. doi: 10.1111/j.1540-8159.1978.tb03466.x.
This report describes some of the various techniques which may be used to demonstrate intermittent and/or incomplete fracture of a pacing lead. Eleven patients treated with an R-wave-inhibited demand pacer showed signs of paroxysmal arrhythmias. Intraoperative recordings of the ventricular electrogram showed the presence of spurious signals mimicking an R-wave. These signals varied from patient to patient but remained morphologically constant when a given technique was used in a given case. The most significant signals were picked up when pacing and recording on the same lead.
本报告描述了一些可用于显示起搏导线间歇性和/或不完全性断裂的不同技术。11例接受R波抑制型按需起搏器治疗的患者出现阵发性心律失常迹象。心室电图的术中记录显示存在模拟R波的伪信号。这些信号在患者之间各不相同,但当在特定病例中使用特定技术时,其形态保持不变。在同一根导线上进行起搏和记录时,能检测到最明显的信号。