Suppr超能文献

人工椎间盘置换术。至少随访3年的初步报告。

Artificial disc replacement. Preliminary report with a 3-year minimum follow-up.

作者信息

Enker P, Steffee A, Mcmillin C, Keppler L, Biscup R, Miller S

机构信息

Cleveland Spine and Arthritis Center, Division of Orthopaedic Surgery, Ohio.

出版信息

Spine (Phila Pa 1976). 1993 Jun 15;18(8):1061-70.

PMID:8367774
Abstract

Artificial disc replacement was performed in six patients with an average age of 55 years and average follow-up of 3.4 years. Four of the six patients had juxtafusion degeneration, one had multilevel disc degeneration, and one patient had isolated disc resorption. The Acroflex disc (Acromed Corporation, Cleveland, OH), which was used in the replacement, is composed of a rubber core vulcanized to two titanium end plates. The latter have superior posts to provide for initial mechanical fixation and porous in-growth surfaces for long-term fixation. Satisfactory results occurred in four of six patients. Poor results occurred in the presence of deformity that resulted in prosthetic failure and isolated disc resorption. There was an average 8 degrees angular and 2.3-mm translational movement and satisfactory in-growth at all interfaces. Design objectives of endurance, biocompatibility, geometry, kinematics, constraint, dynamics, stability, and fail-safe were met; however, this study is preliminary in nature.

摘要

对6例平均年龄55岁、平均随访3.4年的患者进行了人工椎间盘置换。6例患者中有4例为临近融合节段退变,1例为多节段椎间盘退变,1例为孤立性椎间盘吸收。置换所用的Acroflex椎间盘(Acromed公司,克利夫兰,俄亥俄州)由硫化至两个钛制终板的橡胶芯组成。后者具有突出部分以提供初始机械固定,并具有多孔的向内生长表面以实现长期固定。6例患者中有4例取得了满意的结果。在存在导致假体失败和孤立性椎间盘吸收的畸形时,结果较差。所有界面处平均有8度的角位移和2.3毫米的平移运动,且向内生长情况良好。达到了耐久性、生物相容性、几何形状、运动学、约束、动力学、稳定性和故障安全等设计目标;然而,本研究本质上是初步的。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验