Contrast imaging of new multifocal intraocular lens (MIOL) designs with 5 and 7 refractive zones was examined by in vitro measurement of the modulation transfer function (MTF) and by testing contrast sensitivity with Regan 96% and 11% contrast sensitivity charts after "optical implantation" of MIOLs in 20 healthy subjects by means of a new optical device developed by Reiner. The results were compared with those obtained with a monofocal IOL and a diffractive MIOL. We found that the MTF of both refractive MIOLs showed a reduction in contrast at corresponding spatial frequencies compared with the monofocal and with the diffractive IOL, resulting in a Strehl ratio of below 30%. Contrast sensitivity after "optical implantation" was significantly reduced for the refractive 5- and 7-zone models in high contrast (Regan 96%), but for all MIOL models in low contrast (Regan 11%) compared with the monofocal IOL. The refractive 5-zone MIOL and diffractive MIOL showed no statistical difference in contrast sensitivity.
通过体外测量调制传递函数(MTF),以及在20名健康受试者中借助赖纳研发的新型光学装置对多焦点人工晶状体(MIOL)进行“光学植入”后,使用里根96%和11%对比度敏感度图表测试对比度敏感度,对具有5个和7个折射区的新型多焦点人工晶状体设计进行了对比成像研究。将结果与单焦点人工晶状体和衍射型多焦点人工晶状体的结果进行了比较。我们发现,与单焦点人工晶状体和衍射型人工晶状体相比,两种折射型多焦点人工晶状体的MTF在相应空间频率下的对比度均有所降低,导致斯特列尔比低于30%。在高对比度(里根96%)下,折射型5区和7区模型在“光学植入”后的对比度敏感度显著降低,但与单焦点人工晶状体相比,在低对比度(里根11%)下,所有多焦点人工晶状体模型的对比度敏感度均降低。折射型5区多焦点人工晶状体和衍射型多焦点人工晶状体在对比度敏感度方面无统计学差异。