Suppr超能文献

Differences in the calcium concentration of serum and plasma initially and after storage.

作者信息

Lester E, Varghese Z

出版信息

Ann Clin Biochem. 1977 Jan;14(1):39-44. doi: 10.1177/000456327701400107.

Abstract

Calcium concentrations of plasma and serum in glass and plastic containers initially and after storage for up to 8 months at--20 degrees C were compared. The mean initial calcium concentration of the plasma samples was 2.491 nnol/l and this was significantly higher than the mean serum concentration of 2.440 nnol/l. There were no significant differences in the mean calcium concentrations attributable to differences in glass or plastic as materials for storage containers for either serum or plasma. The mean calcium concentration of serum remained stable over the entire storage period, but that of the plasma samples fell significantly between one month and six months, with no further fall between six months and eight months.