Mhatre R, Krull I S
Department of Chemistry, Northeastern University, Boston, Massachusetts 02115.
Anal Chem. 1993 Feb 1;65(3):283-6. doi: 10.1021/ac00051a016.
Gradient ion-exchange chromatography (IEC) was interfaced with a low-angle laser light scattering photometer (LALLS) and ultraviolet (UV) and refractive index (RI) detectors connected in series for on-line determination of the differential refractive index (dn/dc) of proteins and eventually their molecular weights (Mw's). Interfacing of gradient HPLC with a RI detector was made possible by using two isorefractive buffers, which helped generate stable baselines for the LALLS and RI detectors. An optically modified, laser based RI detector was used for determination of dn/dc. On-line determinations of dn/dc required smaller amounts of sample compared to off-line determinations (3-4 mg vs 200 mg). Also, dn/dc could be measured by making one repetitive injection of the sample, if the molar absorptivity of the sample was known. dn/dc values determined on-line were in close agreement with those determined off-line. Additionally, percent recovery of the sample was calculated on-line, and this provided true concentration detected at the UV detector, which was then used for determination of Mw.
梯度离子交换色谱法(IEC)与低角度激光光散射光度计(LALLS)以及串联连接的紫外(UV)和折光率(RI)检测器相连,用于在线测定蛋白质的微分折光率(dn/dc)并最终测定其分子量(Mw)。通过使用两种等折光缓冲液,实现了梯度高效液相色谱与RI检测器的联用,这有助于为LALLS和RI检测器生成稳定的基线。使用了一种经过光学改进的基于激光的RI检测器来测定dn/dc。与离线测定相比,在线测定dn/dc所需的样品量更少(3 - 4毫克对200毫克)。此外,如果已知样品的摩尔吸光系数,通过对样品进行一次重复进样就可以测量dn/dc。在线测定的dn/dc值与离线测定的值非常吻合。此外,在线计算样品的回收率,这提供了在UV检测器处检测到的真实浓度,然后用于Mw的测定。