The use of lower tourniquet inflation pressures in extremity surgery facilitated by curved and wide tourniquets and an integrated cuff inflation system.

作者信息

Pedowitz R A, Gershuni D H, Botte M J, Kuiper S, Rydevik B L, Hargens A R

机构信息

Department of Orthopaedics, University of California, San Diego.

出版信息

Clin Orthop Relat Res. 1993 Feb(287):237-44.

PMID:8448950
Abstract

Use of the lowest possible cuff inflation pressure should minimize the pathogenic effects of compression beneath the pneumatic tourniquet. Curved tourniquets (designed to fit conically shaped limbs) and wider tourniquets were associated with significantly lower arterial occlusion pressure (AOP) than standard, straight tourniquets on the arms and legs of 26 normal volunteers. These tourniquets were used with an integrated tourniquet inflation system in 29 upper-extremity and 31 lower-extremity surgeries. Mean tourniquet inflation pressures of 183.7 mm Hg and 208 mm Hg were used during various surgical procedures of the arm and leg, respectively. Incomplete hemostasis was associated with elevated systolic blood pressure in several cases, but acceptable surgical hemostasis was achieved by incremental increase of the cuff inflation pressure. Curved cuffs, wide cuffs, and an integrated cuff inflation system should facilitate the use of lower tourniquet inflation pressures in extremity surgery.

摘要

使用尽可能低的袖带充气压力应能将气动止血带下压迫的致病影响降至最低。在26名正常志愿者的手臂和腿部,弯曲止血带(设计用于适配圆锥形肢体)和更宽的止血带与显著低于标准直止血带的动脉闭塞压(AOP)相关。在29例上肢手术和31例下肢手术中,这些止血带与一体化止血带充气系统配合使用。在手臂和腿部的各种手术过程中,平均止血带充气压力分别为183.7毫米汞柱和208毫米汞柱。在几例病例中,止血不完全与收缩压升高相关,但通过逐步增加袖带充气压力实现了可接受的手术止血。弯曲袖带、宽袖带和一体化袖带充气系统应有助于在肢体手术中使用较低的止血带充气压力。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验