[0.5T 时的快速自旋回波(Turbo):回波间距和回波数量对图像对比度的影响]

[Turbo(fast) spin echo at 0.5 T: effect of echo distance and echo number on image contrast].

作者信息

Vahlensieck M, Seelos K, Gieseke J, Reiser M

机构信息

Radiologische Klinik, Universität Bonn.

出版信息

Rofo. 1993 Mar;158(3):260-4. doi: 10.1055/s-2008-1032644.

Abstract

Signal intensities, signal-to-noise ratios (S/N), and contrast-to-noise ratios (C/N) in dependence on the echo distance and echo number of the new turbo (fast) spin-echo technique are analysed by phantom, volunteer and patient measurements. With increasing echo number (3 to 15 echoes), signal intensities increase (25 to 50%), S/N varies between different tissues, and C/N of fat to water decreases (70%). A high echo number, although it shortens the imaging time, is thus not always the best choice since the fat/water contrast is lost. However, a high fat/water contrast is advantageous for several MR applications.

摘要

通过模型、志愿者和患者测量,分析了新的涡轮(快速)自旋回波技术中信号强度、信噪比(S/N)和对比噪声比(C/N)与回波间距和回波数量的关系。随着回波数量增加(从3个回波到15个回波),信号强度增加(25%至50%),不同组织间的S/N有所变化,脂肪与水的C/N降低(70%)。因此,尽管高回波数量缩短了成像时间,但由于脂肪/水对比度丧失,它并不总是最佳选择。然而,高脂肪/水对比度对多种磁共振应用有利。