Saksena S, Luceri R, Krol R B, Brownstein S, Burkhardt E, Accorti P, Brewer G, Scott S, Callaghan F, Livingston A
Arrhythmia and Pacemaker Service, Eastern Heart Institute, Passaic, New Jersey 07055.
Am J Cardiol. 1993 Apr 1;71(10):834-41. doi: 10.1016/0002-9149(93)90833-x.
The clinical efficacy and safety of a second-generation braided endocardial pacing, cardioversion and defibrillation lead system was evaluated in 25 patients with ventricular tachycardia (VT) or ventricular fibrillation (VF). The lead system consisted of two 8Fr active fixation endocardial leads each with pacing and defibrillation electrodes and a thoracic patch electrode. Monophasic and biphasic shocks were delivered using a triple-electrode configuration with a right ventricular common cathode and right atrial and thoracic patch anodes. VT and VF were electrically induced. Rapid VT (rate > or = 180 beats/min) and VF were initially terminated by 20 J (550 V) shocks and slow VT (rate < 180 beats/min) by 10 J (400 V) shocks. One hundred fourteen episodes (rapid VT/VF 73, slow VT 41) were treated with 128 shocks (monophasic 80, biphasic 48). Mean ventricular pacing threshold was 0.7 +/- 0.5 ms before and 0.9 +/- 0.5 ms after endocardial shock delivery (p > 0.2). Mean ventricular electrogram amplitude in sinus rhythm was 11.9 +/- 5.7 mV before and 11.4 +/- 5.1 mV after shock delivery (p > 0.2). Simultaneous monophasic endocardial shocks terminated 53% of VF episodes at < or = 20 J. Simultaneous biphasic shocks terminated 94% of all VF episodes at < or = 20 J (p < 0.03). Efficacy of > or = 10 J shocks for rapid VT/VF was greater for biphasic (92%) versus monophasic (74%) shocks (p < 0.05) at lower average shock energy (15 +/- 7 J vs 19 +/- 7 J, respectively, p < 0.05).(ABSTRACT TRUNCATED AT 250 WORDS)
对25例室性心动过速(VT)或室颤(VF)患者评估了第二代编织式心内膜起搏、心脏复律和除颤导线系统的临床疗效及安全性。该导线系统由两根8Fr主动固定心内膜导线组成,每根导线均带有起搏和除颤电极以及一个胸壁贴片电极。采用右心室公共阴极以及右心房和胸壁贴片阳极的三电极配置进行单相和双相电击。通过电刺激诱发VT和VF。快速VT(心率≥180次/分钟)和VF最初通过20J(550V)电击终止,缓慢VT(心率<180次/分钟)通过10J(4o0V)电击终止。对114次发作(快速VT/VF 73次,缓慢VT 41次)进行了128次电击治疗(单相80次,双相48次)。心内膜电击后平均心室起搏阈值在电击前为0.7±0.5ms,电击后为0.9±0.5ms(p>0.2)。窦性心律时平均心室电图振幅在电击前为11.9±5.7mV,电击后为11.4±5.1mV(p>0.2)。单相心内膜电击在≤20J时终止了53%的VF发作。双相电击在≤20J时终止了94%的所有VF发作(p<0.03)。在较低平均电击能量下(分别为15±7J和19±7J,p<0.05),双相电击(92%)对快速VT/VF的≥10J电击疗效高于单相电击(74%)(p<0.05)。(摘要截选至250词)