Suppr超能文献

[美敦力6500型临时心肌起搏导线:心脏手术后三周内的临床评估报告]

[Medtronic model 6500 temporary myocardial pacing lead: a report of clinical assessment until three weeks after cardiac surgery].

作者信息

Asada T, Ogawa K, Mukohara N, Nishiwaki M, Higami T, Sugimoto T, Okada K, Kawamura T

机构信息

Division of Cardiovascular Surgery, Hyogo Brain and Heart Center, Himeji, Japan.

出版信息

Kyobu Geka. 1993 Apr;46(4):327-31.

PMID:8468858
Abstract

Medtronic model 6500 heart wire as a right ventricular electrode was assessed in 41 postoperative cases. R-wave amplitude, slew rate, peak-to-peak, output, current, and resistance were measured by Medtronic A-V pacing system analyser, model 5311, for three weeks after open heart surgery. These parameter changed until the 7 th postoperative day, but did not change significantly thereafter. Output and current did not exceed beyond 10 V and 20 mA respectively, where there is limitations of usually available pacemaker. No complication was encountered. These results showed that this heart wire was reliably useful even until 3 weeks after open heart surgery.

摘要

美敦力6500型心内导线作为右心室电极在41例术后患者中进行了评估。在心脏直视手术后三周内,使用美敦力5311型房室起搏系统分析仪测量R波振幅、上升速率、峰峰值、输出、电流和电阻。这些参数在术后第7天前发生变化,但此后无显著变化。输出和电流分别未超过10V和20mA,这是通常可用起搏器的限制范围。未出现并发症。这些结果表明,即使在心脏直视手术后3周,这种心内导线仍可靠有用。