Yokoyama M, Wada J
J Thorac Cardiovasc Surg. 1978 Mar;75(3):443-5.
A new temporary myocardial electrode has a short piece of needle at one end of an electrode wire. Thus the electrode leaves no exposed areas when implanted. Only the tip is stiched into the myocardium. The device, tested in 20 patients who underwent cardiac operation, has significantly reduced the increase in myocardial threshold to pacing that occurs with the commonly used electrode. This new electrode can be much more easily fixed on both atrial and ventricular myocardium. On postoperative day 7 the threshold measured by Medtronic Model 5880 A was 1.6 mA. on the ventricle and 1.3 mA. on the atrium. The stimulation threshold of the right atrium is almost equal to that of the right ventricle.
一种新型临时心肌电极在电极导线的一端有一小段针。因此,该电极植入时不会留下暴露区域。只有尖端缝入心肌。该装置在20例接受心脏手术的患者中进行了测试,显著降低了与常用电极相比起搏时心肌阈值的升高。这种新型电极可以更容易地固定在心房和心室心肌上。术后第7天,用美敦力5880 A型测量的心室阈值为1.6 mA,心房阈值为1.3 mA。右心房的刺激阈值几乎与右心室相等。