Suppr超能文献

Evaluation of a computerized arrhythmia alarm system.

作者信息

Frost D A, Yanowitz F G, Pryor T A

出版信息

Am J Cardiol. 1977 Apr;39(4):583-7. doi: 10.1016/s0002-9149(77)80169-0.

Abstract

A prospective study of a computerized arrhythmia alarm system was carried out in the coronary care unit during 200 patient hours of monitoring. The computer system was designed to activate an alarm on the development of rhythm and conduction disorders including asystole, ventricular tachycardia, atrial tachycardia, sinus tachtcardia, bradycardia, frequent premature ventricular beats, atrial fibrillation and bundle branch block. Study patients were simultaneously monitored by the computer system and a conventional analog heart rate alarm system. All alarms generated by the two systems were evaluated. Of 79 computer alarms, 42 (53 percent) were true positive alarms; during the same period there were 167 analog alarms of which only 13 (8 percent) were true positive alarms. In both systems, false positive alarms were primarily due to patient movement, but they occurred only 25 percent as often with the computer system as with the analog system. These results indicate that computerized arrhythmia monitoring systems offer significant advantages over conventional monitoring techniques.