Hoenich N A, Smirthwaite P T, Woffindin C, Lancaster P, Frost T H, Vanholder R
Department of Medicine, School of Clinical Medical Sciences, University of Newcastle upon Tyne, UK.
Int J Artif Organs. 1993 Feb;16(2):63-70.
Recirculation is an important factor in single needle dialysis and, if high, can compromise treatment efficiency. To provide information regarding recirculation characteristics of access devices used in single needle dialysis, we have developed a new technique to characterise recirculation and have used this to measure the recirculation of a Terumo 15G fistula needle and a VasCath SC2300 single lumen catheter. The experimentally obtained results agreed well with those established clinically (8.5 +/- 2.4% and 18.4 +/- 3.4%). The experimental results have also demonstrated a dependence on access type, pump speeds and fistula flow rate. A comparison of experimental data with theoretical predictions showed that the latter exceeded those measured with the largest contribution being due to the experimental fistula.
再循环是单针透析中的一个重要因素,如果再循环率较高,会影响治疗效率。为了提供有关单针透析中使用的血管通路装置再循环特性的信息,我们开发了一种表征再循环的新技术,并使用该技术测量了泰尔茂15G动静脉内瘘穿刺针和VasCath SC2300单腔导管的再循环情况。实验获得的结果与临床确定的结果吻合良好(分别为8.5±2.4%和18.4±3.4%)。实验结果还表明再循环率取决于血管通路类型、泵速和内瘘血流量。将实验数据与理论预测值进行比较,结果表明理论预测值超过了实测值,其中最大的差异来自实验用的内瘘。