Rovai D, Lombardi M, Taddei L, Mazzarisi A, Landini L, Ghelardini G, Distante A, Benassi A, L'Abbate A
C.N.R. Clinical Physiology Institute, Pisa, Italy.
Int J Card Imaging. 1993 Mar;9(1):21-7. doi: 10.1007/BF01142929.
The combination of a standardized echographic contrast agent with the analysis of the ultrasonic radio frequency (RF) signal allowed in vitro flow quantitation in a circulation model. The purpose of this study was to investigate both the effects of biological tissues, intervening between probe and insonated structure, and the effects of the angle of incidence between flow and ultrasonic beam on RF flow quantitation. Thus, the contrast agent SHU 454 was intravenously injected (0.4 ml) as a bolus into a circulation model, at variable flow rates, while keeping the pressure and volume of the vessel constant. Injections were performed with saline interposed between probe and vessel and after the addition of the subcutaneous tissue of a pig; injections were also performed using the probe normal to the flow and with an angle of incidence of 45 degrees. Echographic data were recorded by a mechanical sector scanner, capable of sampling the RF signal from a region of interest positioned in the center of the vein. Contrast echo time-intensity curves were generated. As expected, both peak intensity and the area under the curves decreased with intervening tissue (-58 and -70% of baseline values, respectively, p < 0.001). Surprisingly, mean transit time also decreased with intervening tissue (from 1.12 +/- 0.25 seconds with saline, to 0.92 +/- 0.13 seconds with tissue, p < 0.001), thus producing a systematic overestimation of flow (21% on the average). To compensate for signal attenuation, contrast injections were repeated in the presence of tissue after increasing the electronic signal amplification (10 dB), and transit time did not significantly differ from control.(ABSTRACT TRUNCATED AT 250 WORDS)
标准化超声造影剂与超声射频(RF)信号分析相结合,可在循环模型中进行体外血流定量分析。本研究的目的是探究介于探头与被检测结构之间的生物组织的影响,以及血流与超声束之间的入射角对RF血流定量分析的影响。因此,将造影剂SHU 454以0.4 ml团注的形式静脉注射到循环模型中,流速可变,同时保持血管压力和容积恒定。在探头与血管之间插入生理盐水以及添加猪的皮下组织后进行注射;注射时探头垂直于血流以及入射角为45度的情况也进行了注射。超声数据由机械扇形扫描仪记录,该扫描仪能够对位于静脉中心的感兴趣区域的RF信号进行采样。生成了造影回声时间-强度曲线。正如预期的那样,随着中间组织的存在,峰值强度和曲线下面积均下降(分别为基线值的-58%和-70%,p<0.001)。令人惊讶的是,平均通过时间也随着中间组织的存在而减少(从生理盐水时的1.12±0.25秒,降至组织存在时的0.92±0.13秒,p<0.001),从而导致血流系统性高估(平均高估21%)。为补偿信号衰减,在组织存在的情况下增加电子信号放大(10 dB)后重复进行造影剂注射,通过时间与对照无显著差异。(摘要截短于250字)