[背景气流对新生儿呼吸气流和呼吸容积测量准确性的影响]

[Effect of background flow on the accuracy of respiratory flow and respiratory volume measurement in newborn infants].

作者信息

Foitzik B, Schmalisch G, Wauer R R

机构信息

Abteilung Neonatologie, Klinik und Poliklinik für Kinderheilkunde.

出版信息

Biomed Tech (Berl). 1995 Oct;40(10):282-8.

PMID:8527640
Abstract

Measurement of ventilation, in particular in preterm infants, is greatly impaired by equipment dead space, with its significant effect on the ventilatory pattern and gas exchange. For patients of this age, therefore, dead-spacefree methods are needed for long-term measurements. Rebreathing can be avoided if the pneumotachograph (PNT) and face mask are flushed with a continuous background flow. The effect of this on the measurements has not yet been investigated in detail. A measuring system comprising two identical baby PNTs (Jaeger/Germany) permitting a background flow of between 0 and 7 l per min was used. Spontaneous breathing was simulated with a 100 ml calibration syringe employing volumes of 20, 40, 60 and 100 ml (Rudolph/USA) and a frequency of 30 min-1. The measurements were carried out with a T-piece from a respirator circuit, a hand mask (50 ml) and a face chamber having a volume of 850 ml (Siemens-Elema/Sweden). To investigate the dynamic properties of the equipment, we employed flow jumps generated with a magnetic valve (response time < 2 ms) and analysed the responses using Fourier analysis. We were unable to find any significant relationship between the accuracy of volume measurement and tidal volume for any of the measured volumes. An increase in background flow resulted in an underestimation of the volume with an error < 3%. We found no influence of the background flow or type of face mask on the frequency response.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

通气量的测量,尤其是早产儿的通气量测量,会因设备死腔而受到极大影响,这对通气模式和气体交换有显著作用。因此,对于这个年龄段的患者,需要采用无死腔方法进行长期测量。如果用持续的背景气流冲洗呼吸流速计(PNT)和面罩,就可以避免重复呼吸。其对测量的影响尚未进行详细研究。使用了一个测量系统,该系统包括两个相同的婴儿PNT(德国耶格公司),允许背景气流在每分钟0至7升之间。用一个100毫升校准注射器模拟自主呼吸,使用20、40、60和100毫升的容量(美国鲁道夫公司),频率为每分钟30次。测量是通过呼吸回路的T形管、一个手面罩(50毫升)和一个容积为850毫升的面罩室(瑞典西门子 - 埃莱玛公司)进行的。为了研究设备的动态特性,我们利用电磁阀产生流量阶跃(响应时间<2毫秒),并使用傅里叶分析来分析响应。对于任何测量容积,我们都未能发现容积测量精度与潮气量之间存在任何显著关系。背景气流增加会导致容积低估,误差<3%。我们发现背景气流或面罩类型对频率响应没有影响。(摘要截短为250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验