[Results of a study of quality control of diagnostic ultrasound equipment].

作者信息

Kollmann C

机构信息

Institut für Biomed. Technik und Physik, Wien.

出版信息

Ultraschall Med. 1995 Oct;16(5):206-9. doi: 10.1055/s-2007-1003205.

Abstract

AIMS

In this paper results of a quality control programme for B-mode ultrasound equipment used in clinical routine are presented. Over a period of 18 months 5 modern scanners with altogether 19 scanheads were periodically controlled.

METHODS

The methods used for measuring the various performance parameters of an ultrasound equipment followed closely standard methods. All measurements were carried out with commercially available tissue-mimicking phantoms, and documented in a uniform protocol.

RESULTS

Based on the results it was found that 11 of the scanheads did not operate correctly; 2 additional ones were severely defective. Only 6 scanheads were without loss of image quality or function. For the 13 malfunctioning scanheads adjustments, repair or even an exchange was necessary.

CONCLUSION

The study demonstrates that quality control performed periodically is necessary to ensure optimal image quality and safe operation of the equipment.

摘要

目的

本文介绍了一项针对临床常规使用的B型超声设备的质量控制计划的结果。在18个月的时间里,对5台配备共19个探头的现代扫描仪进行了定期检测。

方法

用于测量超声设备各种性能参数的方法严格遵循标准方法。所有测量均使用市售的仿组织体模进行,并记录在统一的方案中。

结果

根据结果发现,11个探头运行不正常;另外2个严重有缺陷。只有6个探头没有图像质量或功能损失。对于13个出现故障的探头,需要进行调整、维修甚至更换。

结论

该研究表明,定期进行质量控制对于确保设备的最佳图像质量和安全运行是必要的。