Effects of initial polarity on defibrillation threshold with biphasic pulses.

作者信息

Natale A, Sra J, Dhala A, Jazayeri M, Deshpande S, Axtell K, Akhtar M

机构信息

Electrophysiology Laboratory, Milwaukee Heart Institute of Sinai Samaritan Medical Center, Wisconsin, USA.

出版信息

Pacing Clin Electrophysiol. 1995 Oct;18(10):1889-93. doi: 10.1111/j.1540-8159.1995.tb03837.x.

Abstract

BACKGROUND

Previous studies have shown that the polarity of epicardial patches significantly affects the defibrillation efficacy of monophasic shocks. However, whether this improvement can be extended to different pulsing methods and lead systems, such as biphasic shocks using endocardial defibrillating electrodes, is unknown.

METHODS

Twenty consecutive patients undergoing testing and permanent implant using an Endotak lead system with a biphasic device were included in the study. In each patient the defibrillation threshold was determined delivering biphasic pulses with the distal coil as the cathode and the proximal coil as the anode during the positive phase and with the polarity reversed. The initial electrode polarity tested was chosen randomly. The defibrillation threshold was defined as the lowest pulse amplitude that effectively terminated ventricular fibrillation induced with 60-Hz alternating current. For each biphasic pulse peak voltage, pulse duration, resistance, and stored energy were recorded.

RESULTS

Of the 20 patients, 12 (60%) had lower defibrillation threshold when the proximal coil was negative, whereas only 2 patients had a lower defibrillation threshold when the distal coil was negative. In four patients a subcutaneous patch would have been required if only the biphasic pulse with the distal coil as negative had been tested. The mean stored defibrillation threshold energy was lower with the configuration using the proximal coil as cathode (16.3 +/- 8.8 J vs 21.5 +/- 11 J; P < 0.01).

CONCLUSION

Change in the initial polarity of biphasic shocks may influence defibrillation efficacy and should, therefore, be assessed in each patient to achieve a more satisfactory safety margin and minimize the use of more invasive lead configurations.