Manhartsberger C, Seidenbusch W
Department of Orthodontics, University Clinic for Dentistry, Innsbruck, Austria.
Am J Orthod Dentofacial Orthop. 1996 Jan;109(1):8-21. doi: 10.1016/s0889-5406(96)70158-0.
Sentalloy springs (GAC, Central Islip, N.Y.) of the open and closed type were investigated with a special designed device. The closed coil springs were subjected to a tensile and the open coil springs to a compression test. After a first measurement, the springs were activated for a period of 4 weeks and then reinvestigated with the same procedure. It could be shown distinctly that, with the different coil springs, the force delivery given by the producer could be achieved only within certain limits. To remain in the martensitic plateau, changed activation ranges, and for the Sentalloy coil springs white and red of the open and closed type, also changed force deliveries had to be taken into account. There was a distinct decrease in force delivery between the first and second measurement. After considering the loading curves of all the Sentalloy coil springs and choosing the right activation range respective to the force delivery, it was found that the coil springs deliver a superior clinical behavior and open new treatment possibilities.
使用一种特殊设计的装置对开放式和封闭式Sentalloy弹簧(GAC,纽约州中央伊斯利普)进行了研究。对封闭式螺旋弹簧进行拉伸试验,对开放式螺旋弹簧进行压缩试验。在首次测量后,弹簧激活4周,然后用相同程序重新研究。可以清楚地表明,对于不同的螺旋弹簧,生产商给出的力传递仅在一定限度内才能实现。为了保持在马氏体平台,改变激活范围,对于开放式和封闭式的白色和红色Sentalloy螺旋弹簧,还必须考虑改变的力传递。第一次和第二次测量之间力传递明显下降。在考虑了所有Sentalloy螺旋弹簧的加载曲线并选择与力传递相应的正确激活范围后,发现螺旋弹簧具有优异的临床性能并开辟了新的治疗可能性。