Miyamoto N, Shimizu Y, Nishiyama G, Mashima S, Okamoto Y
Showa University Fujigaoka Hospital, Yokohama, Japan.
J Electrocardiol. 1995 Oct;28(4):336-7. doi: 10.1016/s0022-0736(05)80054-8.
Body surface potential mapping was performed in 60 clinical cases and an ideal 0 potential was calculated in each case at 2msec intervals, which corresponds to the potential at infinity. Maximal deviation of the Wilson terminal voltage the ideal potential was 0.14mv on the average. Time course of potential variations of the central terminal was in proportion to the measured surface potentials. The lead vector of Wilson terminal was determined for each case, with a method to make a minimal difference between calculate and observed Wilson terminal voltage. The lead vector was directed superiorly and posteriorly with the magnitude of 24% of that of lead I on the average.
对60例临床病例进行了体表电位标测,并每隔2毫秒计算每个病例的理想零电位,该电位对应于无穷远处的电位。威尔逊端电压与理想电位的最大偏差平均为0.14毫伏。中心端电位变化的时间过程与测量的表面电位成比例。通过一种使计算的和观察到的威尔逊端电压之间差异最小的方法,为每个病例确定了威尔逊端的导联向量。导联向量平均向上和向后指向,其大小平均为I导联的24%。