Snider M T, High K M, Richard R B, Panol G, Campbell E A, Service C V, Stene J K, Ultman J S
Department of Anesthesia, College of Medicine, Penn State University, Hershey, USA.
ASAIO J. 1994 Jul-Sep;40(3):M533-9. doi: 10.1097/00002480-199407000-00057.
Blind-ended, hollow fibers mounted on a pulmonary artery catheter may allow O2 and CO2 transfer in the vena cava, right ventricle, and pulmonary artery. The effects of fiber length, manifold number, and gas oscillation on mass and momentum transfer with water perfusate using mass spectrometry and mass flow controllers were studied. Manifolds with 112-196 microporous polypropylene fibers were mounted on 8 Fr multiple lumen, commercially available pulmonary artery catheters. Fiber lengths varied from 0.5 to 16 cm and surface areas from 7 to 220 cm2. Prototypes with 2 cm long fibers were constructed with 1-15 manifolds. A two manifold prototype with 8 cm long fibers and a surface area of 378 cm2 was also studied. The transfer failed to scale with manifold number because the steady gas flow was maldistributed to the manifolds. Oscillating gas pressures from 780 to 76 mmHg absolute at a rate of 40 cycles/min increased CO2 transfer up to 15-fold and O2 transfer up to 2.5-fold. Oscillation also corrected the maldistribution. Optimal fiber lengths of 3 and 1 cm for O2 and CO2, respectively, were seen with steady gas flow, and 8 cm for both with oscillatory gas flow.
安装在肺动脉导管上的盲端空心纤维可能允许氧气和二氧化碳在腔静脉、右心室和肺动脉中进行交换。使用质谱仪和质量流量控制器,研究了纤维长度、歧管数量和气体振荡对水灌注液质量和动量传递的影响。将具有112 - 196根微孔聚丙烯纤维的歧管安装在8F多腔、市售的肺动脉导管上。纤维长度从0.5厘米到16厘米不等,表面积从7平方厘米到220平方厘米不等。构建了纤维长度为2厘米、带有1 - 15个歧管的原型。还研究了一个纤维长度为8厘米、表面积为378平方厘米的双歧管原型。由于稳定气流在歧管间分布不均,传递未能随歧管数量成比例变化。以40次/分钟的速率在780至76毫米汞柱绝对压力之间振荡的气体压力使二氧化碳传递增加了15倍,氧气传递增加了2.5倍。振荡还纠正了分布不均的问题。在稳定气流下,氧气和二氧化碳的最佳纤维长度分别为3厘米和1厘米,在振荡气流下,两者的最佳纤维长度均为8厘米。