[High intensification film-screen systems in thoracic diagnosis. A clinical comparative study].

作者信息

Schäfer C B, Sokiranski R, Claussen C D

机构信息

Abt. für Strahlentherapie CT und Nuklearmedizin KH Barmherzige Brüder, Regensburg.

出版信息

Aktuelle Radiol. 1995 Nov;5(6):389-93.

PMID:8580141
Abstract

OBJECTIVE

The quality of chest images was evaluated for a conventional screen-film system, a new asymmetric screen-film system, and a new uv screen-film system.

MATERIALS AND METHODS

138 chest radiographs (69 p.a., 69 lateral) obtained with three different high intensification screen-film combinations were compared. Film density and film contrast were measured. Three readers graded the image quality according to 16 criteria.

RESULTS

The asymmetric film-screen combination Insight HC showed the lowest film contrast, the best exposure range, and a elevated film density in the mediastinal area. The asymmetric screen-film system was ranked by all three observers as being substantially better in image quality.

CONCLUSION

Compared to conventional screen-film systems, the new screen-film systems can improve the image quality of chest radiographs. Therefore, the new high intensification screen-film combinations can be used as a low cost alternative to the Amber technique and digital radiography.