Doyle K J, Mills D, Larky J, Kessler D, Luxford W M, Schindler R A
Department of Otolaryngology-Head and Neck Surgery, University of California Irvine Medical Center, Orange 92668, USA.
Am J Otol. 1995 Sep;16(5):676-81.
A study was carried out to compare the consonant perceptions of five Nucleus multichannel cochlear implant users, five Clarion-CIS users, and four Clarion-CA users. All subjects were administered the Iowa Medical Consonant Recognition Test in the auditory-only condition. The groups averaged approximately 1 year of implant use at the time of testing, and the average age at implantation was 47 years for the Nucleus users, 60 years for the Clarion-CIS users, and 56 years for the Clarion-CA users. The percent-corrected scores for all three devices were variable and ranged from 34% to 84%, with each group averaging about 50% correct consonant recognition. Consonant feature information transmission was calculated for five features: manner of articulation, nasality, place of articulation, duration, and voicing. The highest information transmission in the Nucleus group was for the features of duration (71%), nasality (60%), and voicing (57%), which agrees with previous studies. The Clarion-CIS group had the highest information transmission for duration, place of articulation, manner of articulation, and nasality (50%, 29%, 28%, and 27%). The Clarion-CA group had the highest information transmission for the features voicing, place of articulation, and duration (41%, 40%, and 37%). This study indicates that the Clarion device may carry some acoustic information related to place of articulation, a feature that has not been found to be encoded well by other multichannel cochlear implants.
开展了一项研究,以比较5名Nucleus多通道人工耳蜗使用者、5名Clarion-CIS使用者和4名Clarion-CA使用者对辅音的感知。所有受试者均在仅听觉条件下接受爱荷华医学辅音识别测试。测试时,各小组平均使用人工耳蜗约1年,Nucleus使用者的平均植入年龄为47岁,Clarion-CIS使用者为60岁,Clarion-CA使用者为56岁。所有三种设备的校正百分比分数各不相同,范围为34%至84%,每组辅音识别正确率平均约为50%。计算了五个特征的辅音特征信息传递:发音方式、鼻音、发音部位、时长和浊音。Nucleus组中,时长(71%)、鼻音(60%)和浊音(57%)特征的信息传递最高,这与之前的研究一致。Clarion-CIS组在时长、发音部位、发音方式和鼻音方面的信息传递最高(分别为50%、29%、28%和27%)。Clarion-CA组在浊音、发音部位和时长特征方面的信息传递最高(分别为41%、40%和37%)。这项研究表明,Clarion设备可能携带一些与发音部位相关的声学信息,这是其他多通道人工耳蜗尚未发现能很好编码的一个特征。