[透析液的电导率。测量、意义以及电解质与临床的相关性]

[Conductivity of the dialysis bath. Measurement, significance and electrolyte and clinical correlations].

作者信息

Ghezzi P, Frigato G, Palanca R

出版信息

Minerva Med. 1977 May 9;68(23):1567-71.

PMID:859709
Abstract

The problem of the conductivity of dialysing solution is discussed. Measurement technique is analysed, highlighting the possibility of error connected with variations in the temperature of the dialysis bath, its mixture, the presence of air bubbles and the presence of dissolved substances that are not electrical charge carriers, and stressing the absolute proportionality with free ions and not with total ions. On the basis of these observations, the lack of any real, reliable correlation between conductivity and electrolytic composition of the dialysing solution unless particular conditions are respected is pointed out. Most important of these are exact knowledge of the composition of the mother solution and the degree of purity of the water. In conclusion, the imprecision of the values given by clinical and technological tradition and widely accepted is pointed out. Such values do not correspond to correct measurement.

摘要

本文讨论了透析液电导率的问题。分析了测量技术,强调了与透析液温度变化、混合情况、气泡存在以及非电荷载流子溶解物质存在相关的误差可能性,并强调了与游离离子而非总离子的绝对比例关系。基于这些观察结果,指出除非遵循特定条件,否则透析液电导率与电解成分之间不存在任何真实、可靠的相关性。其中最重要的是对母液成分和水的纯度的准确了解。总之,指出了临床和技术传统所给出并被广泛接受的值的不精确性。这些值与正确测量不相符。