用于提高腹部磁共振成像信噪比的多次屏气平均法(MBA)

Multiple breath-hold averaging (MBA) method for increased SNR in abdominal MRI.

作者信息

Feinberg D A, Rofsky N M, Johnson G

机构信息

Department of Radiology, New York University Medical Center, New York 10016, USA.

出版信息

Magn Reson Med. 1995 Dec;34(6):905-9. doi: 10.1002/mrm.1910340617.

Abstract

Breath-holding during MR imaging eliminates respiratory motion artifacts but places a major time constraint on data acquisition. This constraint limits image signal-to-noise ratio and hence spatial resolution. A new method, multiple breath-hold averaging, is presented that overcomes these time limitations. Several images are acquired in sequential breath-hold periods, separated by periods of normal breathing, and averaged. This averaged image shows the expected increase in SNR with surprisingly little blurring due to misregistration. SNR improvements can be traded for increased spatial resolution. The MBA methodology can also be applied to 3D data acquisitions, dynamic contrast acquisitions, and image subtractions.

摘要

磁共振成像期间屏气可消除呼吸运动伪影,但对数据采集施加了主要的时间限制。这种限制会降低图像的信噪比,进而影响空间分辨率。本文提出了一种新方法——多次屏气平均法,该方法克服了这些时间限制。在连续的屏气期采集多幅图像,期间穿插正常呼吸期,然后对这些图像进行平均。这幅平均后的图像显示出信噪比预期的提高,且由于配准错误导致的模糊程度小得惊人。信噪比的提高可以用来换取更高的空间分辨率。多次屏气平均法还可应用于三维数据采集、动态对比采集和图像减法。