非气管插管患者使用鼻导管的成本效益分析。

Cost-benefit analysis of nasal cannulae in non-tracheally intubated subjects.

作者信息

Woda R P, Dzwonczyk R, Beckmeyer W, Fuhrman T

机构信息

Department of Anesthesiology, Ohio State University Medical Center, Columbus, USA.

出版信息

Anesth Analg. 1996 Mar;82(3):506-10. doi: 10.1097/00000539-199603000-00014.

Abstract

We evaluated four nasal cannulae used to deliver oxygen and measure PETC02 in a non-tracheally intubated, healthy population. The effect of various oxygen flow rates on PETC02 and respiratory rate (RR), as well as the cost and relative patient comfort of the cannulae, was compared. In this controlled study, 20 healthy volunteers tested the cannulae using oxygen flow rates of 0 (breathing room air), 2, 4, and 6 L/min. The volunteers rated the comfort of each cannula on a scale from 1 (uncomfortable) to 10 (comfortable). Hospital costs of the cannulae were also compared. All of the cannulae, except the Hospitak nasal cannula (HOS), demonstrated an oxygen flow dependency with respect to PETC02 and RR. The HOS cannula was judged by the study population to be the most comfortable and was the second least expensive cannula tested. From a qualitative standpoint, the carbon dioxide waveform recorded with the HOS cannula was the one least affected by oxygen flow. Of the nasal cannula systems evaluated in this study, the HOS system demonstrated the best cost-benefit ratio, performing well clinically while being comfortable to wear and relatively inexpensive. These conclusions are specific to a healthy population and not to patients with lung disease,those who smoke, or those having a higher ASA classification status. Our evaluation suggests that comfort and clinical performance of nasal cannulae may well depend on device design.

摘要

我们评估了四种用于在非气管插管的健康人群中输送氧气并测量呼气末二氧化碳分压(PETC02)的鼻导管。比较了不同氧流量对PETC02和呼吸频率(RR)的影响,以及鼻导管的成本和相对患者舒适度。在这项对照研究中,20名健康志愿者使用0(呼吸室内空气)、2、4和6L/min的氧流量测试了这些鼻导管。志愿者根据1(不舒服)至10(舒服)的量表对每个鼻导管的舒适度进行评分。还比较了鼻导管的医院成本。除了Hospitak鼻导管(HOS)外,所有鼻导管在PETC02和RR方面均表现出氧流量依赖性。研究人群认为HOS鼻导管最舒适,并且是测试的第二便宜的鼻导管。从定性的角度来看,用HOS鼻导管记录的二氧化碳波形受氧流量的影响最小。在本研究评估的鼻导管系统中,HOS系统表现出最佳的成本效益比,在临床上表现良好,佩戴舒适且相对便宜。这些结论仅适用于健康人群,不适用于肺部疾病患者、吸烟者或美国麻醉医师协会(ASA)分级较高的患者。我们的评估表明,鼻导管的舒适度和临床性能很可能取决于设备设计。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验