Anal Chem. 1995 Nov 15;67(22):4217-9. doi: 10.1021/ac00118a028.
The extraction of calcium from water, serum, and urine using Bio-Rex 25 mm ion exchange membrane filters was compared with the oxalate precipitation procedure currently used in our laboratory. Total recoveries of a known quantity of calcium loaded onto the membrane filters for water, serum and urine were as follows: (a) cation exchange filter, 85%, 74%, and 66%; (b) Chelex, 65%, 98%, and 20%; and (c) oxalate precipitation, 93%, 100%, and 96%, respectively. Regression analysis for precipitation versus ion exchange isotope ratio measurements of standards prepared using highly enriched calcium-44 showed slopes of unity. An improvement of automated sample analysis was observed for water and urine calcium samples extracted with ion exchange filters.