Fink U, Schwerdtfeger J, Fink B K, Schätzl M, Reiser M
Institut für Radiologische Diagnostik, Klinikum Grosshadern, Ludwig-Maximilians-Universität München.
Rofo. 1996 Apr;164(4):275-80. doi: 10.1055/s-2007-1015655.
The accuracy of digital luminescence radiography was compared with that of conventional film/screen techniques, using animal preparations and clinical examinations.
Fine fissures were made in 8 animal bones and these were examined radiologically. The digital examinations were carried out with and without edge enhancement. 208 patients were examined in a similar way. Film quality and assessment of the fractures were evaluated quantitatively.
In no instance did either of the digital methods provide inferior quality when compared with conventional films. ROC analysis for evaluation of fractures in patients, using an experienced radiologist, showed no significant difference between the various methods (ROC areas: conventional 0.947, digital 0.958, digital with edge enhancement 0.943). With a less experienced observer there were significant advantages for both digital methods (ROC areas: 0.851, 0.886, 0.908).
Our investigation has proved that fractures which are difficult to see can be reliably demonstrated by digital luminescence radiography.