Neuss M B, Coe J Y, Tio F, Le T P, Grabitz R, Redel D A
Department of Cardiology, Children's Hospital, University of Bonn, Adenauerallee 119, D-5300, Bonn 1, Germany.
Cardiovasc Intervent Radiol. 1996 May-Jun;19(3):170-5. doi: 10.1007/BF02577614.
To examine the feasibility of transvenous placement of a new memory-shaped, small, retrievable coil that has a smaller-caliber delivery system than currently available devices, for closure of the patent ductus arteriosus (PDA).
Through 4 or 5 Fr vascular sheaths and 4 or 5 Fr end-hold catheters, the coils were delivered and placed in piglets (n = 10) with PDA. The coils were made from 0.018" (0.46 mm) or 0.028" (0.71 mm) stainless steel guidewire. Mounted for delivery, the new device has the appearance of a conventional guidewire. This neonatal PDA model was created without major surgery or drugs by stenting the ductus arteriosus.
The memory-shaped coils were easily delivered. Coils not optimally placed were retrieved and repositioned. Occlusion of the ductus arteriosus as early as a half-hour after delivery was shown angiographically and confirmed by histopathology.
This new, small-caliber, simple device was found to be effective for closure of the PDA in this animal model. Longer-term observations are needed.
研究经静脉置入一种新型记忆形状、小型、可回收线圈的可行性,该线圈具有比现有设备更小口径的输送系统,用于闭合动脉导管未闭(PDA)。
通过4或5 Fr血管鞘和4或5 Fr端持导管,将线圈输送并放置于患有PDA的仔猪(n = 10)体内。线圈由0.018英寸(0.46毫米)或0.028英寸(0.71毫米)不锈钢导丝制成。新装置在输送时安装好,外观类似传统导丝。该新生儿PDA模型通过对动脉导管进行支架置入而创建,无需进行大手术或使用药物。
记忆形状的线圈易于输送。未最佳放置的线圈可被回收并重新定位。血管造影显示在输送后半小时内动脉导管即被闭塞,组织病理学证实了这一点。
在该动物模型中,这种新型小口径、简单的装置被证明对闭合PDA有效。需要进行长期观察。