McCreesh Z, Evans N E, Scanlon W G
Northern Ireland Bioengineering Centre, UK.
Med Eng Phys. 1996 Mar;18(2):110-4. doi: 10.1016/1350-4533(95)00037-2.
User-induced errors are common when women repetitively employ conventional probe type thermometers to chart their basal body temperatures in an effort to indicate ovulation. An alternative technique employing a two-part telemetric thermometer is proposed, with low-power, SAWR-controlled UHF radio as the transmission medium. Worn overnight in the vagina, the 1 microW erp telemetry transmitter sends pulse modulated data continuously to a microcontroller in a nearby receiver; a real time clock enables programmable sampling and storage of the subject's temperature to 0.1 degrees C resolution. Initial clinical results indicate an enhanced performance compared to oral and axillary temperature trends taken by a mercury-in-glass thermometer. Polar plots of both the isolated and body-worn telemetry transmitter are presented; body induced attenuations of up to 30 dB were measured.
当女性为了确定排卵情况而反复使用传统探头式温度计来记录基础体温时,人为误差很常见。本文提出了一种采用两部分遥测温度计的替代技术,以低功率、声表面波振荡器控制的超高频无线电作为传输介质。这种1微瓦等效辐射功率的遥测发射器在阴道内佩戴过夜,将脉冲调制数据持续发送到附近接收器中的微控制器;实时时钟可实现对受试者体温的可编程采样和存储,分辨率可达0.1摄氏度。初步临床结果表明,与玻璃水银温度计测量的口腔和腋下体温趋势相比,该技术性能有所提高。文中给出了单独的遥测发射器和佩戴在人体上的遥测发射器的极坐标图;测量到人体引起的衰减高达30分贝。