Biosensor based on chemiluminescence for serum uric acid determination.

作者信息

Liu J, Guo J, Li G

机构信息

Institute of Microbiology, Chinese Academy of Sciences, Beijing, China.

出版信息

Chin J Biotechnol. 1995;11(3):177-83.

PMID:8679934
Abstract

A biosensor based on luminol chemiluminescent reaction and immobilized uricase column for serum uric acid determination has been developed. The response time by the sensor for various concentrations of serum uric acid was 47 sec with a 17-microliters sample volume at 40 samples per hr. The linear range of standard curve was from 1 mg/dL up to 20 mg/dL of serum uric acid. The imprecision within a day was 3.22% to 4.36%. The day-to-day imprecision was 6.18% to 7.80%. The recovery rate by the method was 93% to 109%. Compared with the standard colorimetric method employed, the enzymatic kit revealed that the linear regression and correlation coefficient were Y = 0.4 + 0.938x and r = 0.9909, respectively. The immobilized uricase column retained 94% of its original activity even after over 2000 runs for five and half months of continual usage.

摘要

一种基于鲁米诺化学发光反应和固定化尿酸酶柱的用于测定血清尿酸的生物传感器已被开发出来。该传感器对不同浓度血清尿酸的响应时间为47秒,样品体积为17微升,每小时可检测40个样品。标准曲线的线性范围为血清尿酸1mg/dL至20mg/dL。日内不精密度为3.22%至4.36%。日间不精密度为6.18%至7.80%。该方法的回收率为93%至109%。与采用的标准比色法相比,酶试剂盒显示线性回归和相关系数分别为Y = 0.4 + 0.938x和r = 0.9909。即使经过连续五个半月超过2000次运行,固定化尿酸酶柱仍保留其原始活性的94%。