Zhang Yan, Wen Guangming, Zhou Yehong, Shuang Shaomin, Dong Chuan, Choi Martin M F
Institute of Advanced Chemistry, School of Chemistry and Chemical Engineering, Shanxi University, Taiyuan 030006, China.
Biosens Bioelectron. 2007 Mar 15;22(8):1791-7. doi: 10.1016/j.bios.2006.08.038. Epub 2006 Oct 4.
An uric acid biosensor fabricated from a uricase-immobilized eggshell membrane and an oxygen electrode was presented. The detection schemes involve the enzymatic reactions of the uricase leading to the depletion of dissolved oxygen level upon exposure to uric acid solution. The decrease in oxygen level was monitored and related to the uric acid concentration. The scanning electron micrographs show the microstructure of the eggshell membrane within which the uricase is successfully immobilized. The effects of enzyme loading, pH, temperature, and phosphate buffer concentration on the response of the biosensor were investigated in detail. The uric acid biosensor has a linear response range of 4.0-640 microM with a detection limit of 2.0 microM (S/N=3). The response time was less than 100 s. The biosensor exhibited good repeatable response to a 0.10mM uric acid solution with a relative standard deviation of 3.1% (n=7). The reproducibility of fabrication of the biosensors using four different membranes was good with a R.S.D. of 3.2%. The biosensor showed extremely good stability with a shelf-life of at least 3 months. Some common potential interferents in samples such as glucose, urea, ascorbic acid, lactic acid, glycine, DL-alpha-alanine, DL-cysteine, KCl, NaCl, CaCl2, MgSO4, and NH4Cl showed no interferences on the response of the uric acid biosensor. The biosensor was successfully applied to determine the uric acid level in some human serum and urine samples, and the results agreed well with those obtained by a commercial colorimetric assay kit.
介绍了一种由固定有尿酸酶的蛋壳膜和氧电极制成的尿酸生物传感器。检测方案涉及尿酸酶的酶促反应,该反应在暴露于尿酸溶液时会导致溶解氧水平降低。监测氧水平的降低并将其与尿酸浓度相关联。扫描电子显微镜照片显示了成功固定尿酸酶的蛋壳膜的微观结构。详细研究了酶负载量、pH值、温度和磷酸盐缓冲液浓度对生物传感器响应的影响。该尿酸生物传感器的线性响应范围为4.0 - 640微摩尔,检测限为2.0微摩尔(信噪比 = 3)。响应时间小于100秒。该生物传感器对0.10毫摩尔的尿酸溶液表现出良好的可重复响应,相对标准偏差为3.1%(n = 7)。使用四种不同膜制备生物传感器的重现性良好,相对标准偏差为3.2%。该生物传感器具有极佳的稳定性,保质期至少为3个月。样品中的一些常见潜在干扰物,如葡萄糖、尿素、抗坏血酸、乳酸、甘氨酸、DL-α-丙氨酸、DL-半胱氨酸、氯化钾、氯化钠、氯化钙、硫酸镁和氯化铵,对尿酸生物传感器的响应均无干扰。该生物传感器已成功应用于测定一些人血清和尿液样本中的尿酸水平,结果与商用比色测定试剂盒获得的结果吻合良好。