自动植入式心脏复律除颤器贴片移位导致大量血胸。

Migration of an automatic implantable cardioverter-defibrillator patch causing massive hemothorax.

作者信息

Quigley R L, Hamer M E, Swiryn S

机构信息

Department of Surgery, Northwestern University Medical School, Evanston Hospital, Illinois 60201, USA.

出版信息

Tex Heart Inst J. 1996;23(1):54-7.

Abstract

The automatic implantable cardioverter-defibrillator is composed of shocking electrodes (patches or coils), sensing electrodes, and a generator implanted in the abdominal wall. A number of complications with defibrillators have been reported, including migration of the various components. We describe a patient in whom one of the patches, originally sutured extrapericardially, migrated in the right hemithorax and eroded into the ipsilateral lung. The resultant massive hemothorax necessitated urgent thoracotomy, patch removal, and hematoma evacuation. A hybrid implantable cardioverter-defibrillator system was created as a replacement to minimize surgery while protecting the patient from sudden death.

摘要

植入式自动心脏除颤器由电击电极(贴片或线圈)、传感电极和植入腹壁的发生器组成。已经报道了许多除颤器相关的并发症,包括各个组件的移位。我们描述了一名患者,其最初心包外缝合的一个贴片移位至右半胸并侵蚀到同侧肺脏。由此导致的大量血胸需要紧急开胸手术、移除贴片和清除血肿。创建了一种混合植入式心脏除颤器系统作为替代方案,以尽量减少手术,同时保护患者免于猝死。