Department of Cardiology, Manchester Heart Centre, Manchester Royal Infirmary, Manchester, UK.
Can J Cardiol. 2013 Aug;29(8):1015.e9-1015.e10. doi: 10.1016/j.cjca.2012.12.006. Epub 2013 Apr 15.
In recent years the phenomenon of reverse twiddler's syndrome has been described, characterized by pulse generator manipulation resulting in lead advancement rather than retraction. We describe what we believe to be the first reported case of both classic and reverse twiddler's syndrome occurring simultaneously in a patient with a biventricular implantable cardioverter-defibrillator.
近年来,出现了一种被称为“反向扭动器综合征”的现象,其特征是脉冲发生器的操作导致导线前进而不是回缩。我们描述了一例患者同时出现经典型和反向扭动器综合征的情况,该患者装有双心室植入式心脏复律除颤器,我们认为这是首例报告。