Dacron-covered stent-grafts for the percutaneous treatment of carotid aneurysms: effectiveness and biocompatibility--experimental study in swine.

作者信息

Link J, Feyerabend B, Grabener M, Linstedt U, Brossmann J, Thomsen H, Heller M

机构信息

Department of Radiology, Christian-Albrechts-University of Kiel, Germany.

出版信息

Radiology. 1996 Aug;200(2):397-401. doi: 10.1148/radiology.200.2.8685332.

Abstract

PURPOSE

To evaluate the short-term effectiveness and biocompatibility of Dacron-covered stent-grafts for percutaneous treatment of carotid aneurysms.

MATERIALS AND METHODS

In eight swine, nine aneurysms were created surgically in the common carotid artery. Percutaneous treatment was performed with Dacron-covered stent-grafts. Seven of eight swine underwent follow-up angiography at 2, 3, 4, and 5 weeks after stent implantation. Histologic examinations of tissue from the site of stent implantation were performed at the same intervals.

RESULTS

Initial closing off of the aneurysm was possible in all swine. Angiographic follow-up revealed patency of only one Dacron-covered stent-graft after 2 weeks. In six of seven swine, the stent was occluded on the follow-up angiogram. One of the swine died of unclear cause after 1 week. The stent-graft of this swine was not occluded. In all Dacron-covered stent-grafts, inflammatory reaction with granulocyte infiltration was found next to the Dacron material.

CONCLUSION

Primary treatment of carotid aneurysms with Dacron-covered stent-grafts is effective. However, the short-term patency rate is poor, which probably can be attributed to the limited biocompatibility of the Dacron cover. An acute inflammatory reaction against Dacron seems responsible for the poor patency rate.