眼部眼轴长度测量的超声法与摄影法比较。

Comparison of ultrasonic and photographic methods of axial length measurements of the eye.

作者信息

Grey R H, Perkins E S, Restori M

出版信息

Br J Ophthalmol. 1977 Jun;61(6):423-7. doi: 10.1136/bjo.61.6.423.

Abstract

The axial length of 25 eyes was measured by the photographic method and by ultrasonography. The photographic measurements were consistently larger than the ultrasonic, but analysis gave a correlation coefficient of r = 0-8483. The results suggest that the centre of rotation of the eye lies posterior to the anatomical centre of the globe. Using an empirical method of calculation and allowing for an error of +/-0-3 mm in the ultrasonic measurement, the photographic results were within +/-0-5 mm in 13 eyes and +/-1-0 mm in 7 eyes. The photographic method may be helpful as a guide to the power of intraocular lens implants and for comparing axial lengths in population studies.

摘要

采用摄影法和超声检查法测量了25只眼睛的眼轴长度。摄影测量结果始终大于超声测量结果,但分析得出相关系数r = 0.8483。结果表明,眼球的旋转中心位于眼球解剖中心的后方。采用经验计算方法,并考虑到超声测量中±0.3 mm的误差,13只眼睛的摄影测量结果在±0.5 mm以内,7只眼睛在±1.0 mm以内。摄影法可能有助于指导人工晶状体植入的屈光度,并用于人群研究中眼轴长度的比较。

相似文献

5
Simple method of determining the axial length of the eye.测量眼轴长度的简易方法。
Br J Ophthalmol. 1976 Apr;60(4):266-70. doi: 10.1136/bjo.60.4.266.
10
[The measurement of eye axial length by ultrasound].[超声测量眼轴长度]
Zhongguo Yi Liao Qi Xie Za Zhi. 1997 Jan;21(1):24-5.

本文引用的文献

1
Ophthalmic biometry using ultrasound.超声眼科生物测量法
Int Ophthalmol Clin. 1969 Fall;9(3):667-83. doi: 10.1097/00004397-196900930-00009.
2
Simple method of determining the axial length of the eye.测量眼轴长度的简易方法。
Br J Ophthalmol. 1976 Apr;60(4):266-70. doi: 10.1136/bjo.60.4.266.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验