This paper investigates the effect of lubricant on the size and morphology of ultra-high molecular weight polyethylene (UHMWPE) wear debris produced by a hip joint simulator. Bovine serum and water were chosen as the lubricants for comparison. When bovine serum was used as lubricant, the majority of the wear debris produced was submicron sized and equiaxed or elongated in shape; no transfer film was found on the femoral heads. When water was used as lubricant, large pieces of thin UHMWPE flakes about 2-3 mm across were produced and patches of transfer films were observed on the femoral heads. These large flakes were comprised of numerous smaller thin flakes compacted together. While the size and shape of the wear debris produced with serum lubrication are similar to those reported from clinical studies in total hip replacement, the debris produced with water lubrication bears no resemblance to what has been seen clinically.